Bio-Ray SDX
Digital X-Ray Sytem
Visualize Quality
Bio-Ray SDX size 2 sensor
Needle Structure Scintillator:
The purpose of the scintillator is to convert x-rays to light. A needle
structure scintillator helps to focus the photons in a straight, downward
fashion, eliminating unwanted scatter. Most companies utilize a granular
structure scintillator, which generates scatter and unwanted noise.
Fiber Optic Plate:
The fiber optic plate improves image quality by stopping nonconverted
x-rays from interacting directly with the CCD, which in return eliminates
unwanted noise in the signal. It also helps to focus the photons and
apply them to the CCD in an even fashion.
CCD:
A unique manufacturing process enables us to provide a CCD with little
or no blemishes. This almost flawless CCD requires no correction file,
which drastically improves hardware performance. A correction file is
software that is needed to fix defects in the hardware. With no correction
file your sensor will be ready instantly, and more importantly you won’t
have to worry about your sensor timing out during the capture process.
Eliminating the need for a correction file also improves workflow with
multiple sensors.
Bio-Ray SDX:
By utilizing a unique manufacturing process combined with the ideal
materials for dental CCD imaging, and one of the smallest pixel sizes
available, Bio-Ray SDX offers superior resolution and image quality.
Sigma Digital X-ray combines superior hardware, software and service,
making the Bio-Ray SDX the best Intraoral Digital Radiography System
available today.
Schematic of a Scintillator-CCD Detector
Schematic of a Scintillator-CCD Detector. Incident X-ray photons interact
with the scintillator where they are absorbed and light photons are
emitted. Light photonsenter the fiber optic plate and are carriedto
the CCD where they are detected and converted to electrons.
Comparison between a needle and granular
scintillator material
These illustrations show the emission spectra, decay, and X-ray absorption
coefficient characteristics between the needle and granular scintillator
materials. The Bio-Ray SDX uses a needle structure scintillator, as
seen here on the left. This comparison clearly shows how the needle
structure scintillator eliminates unwanted scatter.
Intraoral Sensor Specifications: |
| Sensor Size: |
|
Size 1 |
Size 2 |
| Pixels in Sensor |
|
1.50 |
2.04 |
| Pixel Size: |
|
20 x 20 |
20 x 20 |
| Active Image Area(mm): |
|
30 x 20 |
34 x 24 |
| Resolution: |
|
25 |
25 |
| External Dimensions(mm) |
|
38.5 x 26.2 |
42.5 x 30.5 |
Bio-Ray Intraoral Digital Radiography System Includes: